Olympus was informed that two patients developed klebsiella pneumonia infections after undergoing cystoscopy procedures using two of the facility¿s cystonephrofiberscopes (sn.(b)(4), sn.(b)(4)).The patients were treated with intravenous antibiotics.In addition, the user facility reported that they use enzyme and aldahol cleaner to disinfect and clean the scope.The minimum effective concentration is being checked every 2 weeks.The endoscope channel is being brushed during manual cleaning using a reusable brush (model: unk).The scope is precleaned immediately after the procedure as it is rinsed, flushed, soaped and the steps are then repeated.The scope is not being leaked tested prior to manual cleaning.Air is flushed into the scope after reprocessing.There is no known issue with the facility¿s automatic endoscope reprocessor (aer) (model: unk).The facility's last in-service with an olympus endoscopy support specialist was conducted on (b)(4) 2019.There have been no changes to the facility¿s reprocessing staff since the last in-service.The user facility reported that all the reprocessing staffs were trained on how to properly reprocess an endoscope.This is 1 of 2 reports.
|
The scope was sent to an independent laboratory for microbial testing.The scope's instrument channel was cultured and no klebsiella organism or any other microbial growth was recovered from the scope.The scope was then ethylene oxide (eto) sterilized and returned to olympus for a device evaluation.Olympus performed a visual inspection on the scope using an olympus boroscope and a telescope which found discoloration on the wall of biopsy channel at the distal end opening, kink near the middle of channel, and slight brownish stain on the k-mount near the channel opening on the control body side.There was also chemical damage/dried residue noted on the light guide lens, bending section cover, bending section cover glue and insertion tube.The scope passed leak testing.Based on the evaluation findings, the kink within the channel is due to mishandling.The cause of the discoloration/chemical damage on the bending section cover, bending section cover glue, insertion tube, and biopsy channel (brownish stain) is due to improper reprocessing.The instruction manual provides the users several warnings to ¿after using this instrument, reprocess and store it according to the instructions given in the endoscope¿s companion reprocessing manual.Using improperly or incompletely reprocessed or stored instruments may cause patient cross-contamination and/or infection.Do not strike, bend, hit, pull, twist, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope with excessive force.The endoscope may be damaged.¿.
|