• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY ECHOTIP ULTRA FIDUCIAL NEEDLE; NEU, MARKER, RADIOGRAPHIC, IMPLANTABLE, BIOPSY NEEDLE KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY ECHOTIP ULTRA FIDUCIAL NEEDLE; NEU, MARKER, RADIOGRAPHIC, IMPLANTABLE, BIOPSY NEEDLE KIT Back to Search Results
Model Number G03793
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/31/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Prior to distribution, all echotip ultra fiducial needles are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic ultrasonography (eus), the physician used a cook echotip ultra fiducial needle.After the deployment of the third fiducial marker, the needle was unable to retract to point zero (0) in length.The fourth fiducial marker could not be deployed and remained stuck in the needle.This caused the deployment of three (3) fiducial markers only.The needle was cautiously removed from the endoscope.Other than the deployed fiducials, a section of the device did not remain in the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP ULTRA FIDUCIAL NEEDLE
Common Device Name
NEU, MARKER, RADIOGRAPHIC, IMPLANTABLE, BIOPSY NEEDLE KIT
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key8367141
Report Number1037905-2019-00090
Device Sequence Number1620853
Product Code NEU
UDI-Device Identifier00827002037939
UDI-Public(01)00827002037939(17)190822(10)W4107329
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K141356
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2019
Device Model NumberG03793
Device Catalogue NumberECHO-22-F
Device Lot NumberW4107329
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2019
Initial Report FDA Received Date02/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient Weight68
-
-