• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 3.0NEO
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code Extubate (2402)
Date of Event 09/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the change of tracheostomy tube, the insertion was very painful due to the thick and rigid material.It was also stated that the device needed more pressure to be applied which may also be the reason of the problem.The patient was re-intubated.After a third attempt to introduce the 3rd cannula, a cannula of a different lot number was used without issue.In subsequent insertion of 2 cannulas of the reported lot number, the painfulness and suffocation status were repeated.Also, by checking and by touch control she could feel the difference of the material.The patient has had a permanent tracheostomy for more than 2 years that requires very often changes of tubes.The usage of the tracheostomy tubes is on a regular basis.
 
Additional Manufacturer Narrative
Evaluation summary: a third sample was received for evaluation, no reported event was observed since met with the product specifications and no corrective actions are required.A visual inspection was performed and it was observed all the components are assembled properly at the tube.A dimensional test was performed outer and inner diameter of the cannula this reading is within specification.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
MDR Report Key8367313
Report Number2936999-2019-00136
Device Sequence Number1620881
Product Code JOH
UDI-Device Identifier30884522026653
UDI-Public30884522026653
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
K945513
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/23/2020
Device Model Number3.0NEO
Device Catalogue Number3.0NEO
Device Lot Number15B0657JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2019
Initial Date Received by Manufacturer 12/07/2018
Supplement Date Received by Manufacturer06/13/2019
Initial Report FDA Received Date02/25/2019
Supplement Report FDA Received Date06/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age4 YR
-
-