MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PROMAX¿ SERIES BARD® C-MAX¿ CUTTING LOOP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PROMAX¿ SERIES BARD® C-MAX¿ CUTTING LOOP Back to Search Results
Model Number 355301
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 01/31/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the cutting loop broke during use.The complainant reported that the case was continued and nothing was left in the patient.
 
Additional Manufacturer Narrative
The reported event was confirmed, but the cause was unknown.The evaluation of the returned sample, performed by omnitech systems, stated that all components were present except for the loop.The remaining portion of the loop exhibits an opposite bend to manufacturing angle.Omnitech systems further stated that after a device history record review was performed, one ncr was found for standard 100% inspection of tungsten, in which 60 of 116 pieces were scrapped.There were no other ncr's associated.Omnitech systems concluded that due to the evidence of the opposing bend, this event occurred post-manufacturing.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "potential complications: complications could include local and/or systemic infection, thermal damage to surrounding structures, local hematoma, dissection and perforation, distal tip separation, and patient discomfort during and/or after energy application.Precautions: 1.All medical staff should carefully review product labeling and instruction sheets before using the bard® c-max¿ cutting loop device.Inappropriate use of the instrument could adversely affect the procedure or cause injury to the patient.2.Refer to the applicable operating and maintenance manuals for the resectoscope and electrosurgical generator being used.3.The bard® c-max¿ cutting loop device should be used only by a physician who is familiar with the use of electrosurgical instruments, devices and power generators.Consult the medical literature regarding techniques, typical power settings, complications, and hazards prior to any endoscopic procedure.4.During application of electrical energy, movement of the tip is required to achieve the desired tissue effect.5.The time and energy required for treating specific tissue may differ when using the bard® c-max¿ cutting loop device compared to other electrosurgical devices.6.The monopolar electrosurgical generator should be set to the cut or pure cut mode for maximum tissue removal effect.Do not use the blend mode and be cautious of arcing on high coagulation power settings.7.Starting at power settings consistent with standard resectoscope cutting loop procedures, gradually increase power until the desired tissue effect is achieved.If there is little or no tissue effect, check the generator, power and grounding cables and the electrodes.Various electrosurgical generators have different power delivery capabilities.Depending on the impedance in the monopolar circuit, the power delivered may be lower than the power setting indicates (see owner/operator manual for the generator being used).8.Constant irrigation is required throughout the procedure, and the distal tip of the device should be submerged and kept in view at all times.Use sterile non-conductive irrigation solution only.9.Immediately discontinue use if breaks or fractures appear in the bard® c-max¿ cutting loop device.Breaks or fractures may allow undirected emission of electrical energy, rendering the device useless and potentially causing harm to surrounding tissues.10.Do not bend or manipulate the device.11.Care should be taken to avoid severe impacts, side stresses or bends at sharp angles.12.When endoscopic devices are used together, ensure that any isolation or ground is not violated.Directions for use: 1.Remove the bard® c-max¿ cutting loop device from the package and examine it for damage.Do not use if there are visible signs of damage or insulation is not intact.2.Assemble the working element and resectoscope according to the instructions provided by the resectoscope manufacturer.3.Insert the bard® c-max¿ cutting loop device into the working channel of the resectoscope per instructions indicated in the resectoscope operating manual.4.Ensure that the electrode is securely placed and locked into position by pulling carefully on the electrode stabilizer sleeve (if appropriate).5.Attach resectoscope to the electrosurgical generator according to manufacturer's recommendations.6.Insert the combined assembly into the resectoscope sheath and position at the point where initial application of electrical energy will be delivered.7.Unless recommended by other sources, the electrosurgical generator should be set on the pure cut setting at power levels consistent with standard resectoscope cutting loop procedures.8.The power level can be adjusted and varied slightly for different patients and also when tissue is being removed on subsequent sweeps.9.The level of effect is governed by the amount of time the electrode is in contact with an area and the power delivered.10.The slower the electrode is moved across any particular area and the higher the power applied, the deeper the level of effect.11.Maintain constant irrigation through the resectoscope using sterile non-conductive irrigant.Keep the tip in the field of view and submerged in irrigant at all times.12.If tissue adheres to the tip, cleaning can be accomplished by switching to a coagulation setting and engaging current without tissue contact and with tip fully submerged in irrigation solution.Cleaning can also be accomplished by removing the device and wiping it with a soft, sterile lint-free cloth.13.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations." section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the cutting loop broke during use.The complainant reported that the case was continued and nothing was left in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMAX¿ SERIES BARD® C-MAX¿ CUTTING LOOP
Common Device Name
CUTTING LOOP
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8367364
Report Number1018233-2019-01018
Device Sequence Number18939989
Product Code FAS
UDI-Device Identifier00801741018510
UDI-Public(01)00801741018510
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/24/2022
Device Model Number355301
Device Catalogue Number355301
Device Lot NumberNGCN0345
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2019
Initial Date Received by Manufacturer 02/04/2019
Supplement Date Received by Manufacturer06/03/2019
Initial Report FDA Received Date02/25/2019
Supplement Report FDA Received Date06/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-