| Model Number |
MMT-751RNAP |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had button error.The customer's blood glucose level was unknown at the time of incident.Customer stated that the insulin pump is alarming.Advised the insulin pump will need to be replaced and to discontinue use of the insulin pump and revert to a back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No button error alarm noted during testing.Device had unresponsive buttons due to flattened (creased) all buttons dome switch.No unlocked lcd keypad connector noted.
|
| |
|