• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD ECLIPSE¿ HYPODERMIC NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD ECLIPSE¿ HYPODERMIC NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Catalog Number 305763
Health Effect - Clinical Codes Exposure to Body Fluids (1745); Needle Stick/Puncture (2462)
Date of Event 01/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that a nurse sustained a dirty needle stick injury from the bd eclipse¿ hypodermic needle during an "emergency im injection to an agitated resident".A "portion of the plastic cover" failed to shield part of the needle once activated, resulting in a prick to the nurse's hand.However, no medical intervention was reported.
 
Event or Problem Description
It was reported that a nurse sustained a dirty needle stick injury from the bd eclipse¿ hypodermic needle during an "emergency im injection to an agitated resident".A "portion of the plastic cover" failed to shield part of the needle once activated, resulting in a prick to the nurse's hand.However, no medical intervention was reported.
 
Additional Manufacturer Narrative
Investigation: 100 representative samples in a shelf carton were received for evaluation.All 100 samples were subjected to visual inspection.There were no deformation/damage observed on the eclipse hub and safety shield.No breakage observed on the safety lock pin.10 samples were sent for ct scan for roundness and rod od measurements.The roundness and rod od results met the specifications requirement.20 samples were manually activated in accordance with the instruction for use (ifu) with no abnormalities observed.10 samples were activated per test method procedure, it377.The activation force results met the specification requirements.The complaint is unconfirmed as no defect is observed on the sample.No assignable root cause could be established as the investigation revealed no abnormalities associated with the returned representative samples and all tested samples have passed the inspection.A dhr was performed and no qn's were observed relating to the reported defect on any of the batches.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ECLIPSE¿ HYPODERMIC NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
MDR Report Key8370354
Report Number8041187-2019-00169
Device Sequence Number16418385
Product Code FMI
UDI-Device Identifier30382903057635
UDI-Public30382903057635
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Catalogue Number305763
Device Lot Number8207153
Initial Date Received by Manufacturer 02/11/2019
Supplement Date Received by Manufacturer02/11/2019
Initial Report FDA Received Date02/26/2019
Supplement Report FDA Received Date04/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-