• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 DEPTH GAUGE; PROSTH, HIP, SEMI-CONST, MET/CERAM/POLY, CEMENT OR NON-POROUS, UNCEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 DEPTH GAUGE; PROSTH, HIP, SEMI-CONST, MET/CERAM/POLY, CEMENT OR NON-POROUS, UNCEMENT Back to Search Results
Catalog Number 71364451
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/04/2019
Event Type  malfunction  
Event Description
Tip of depth gauge broke while attempting to measure screw length.No backup was available.No delay to procedure recorded.
 
Manufacturer Narrative
The associated complaint device was returned and evaluated.A visual inspection of the returned r3 depth gauge indicated the tip fractured off, confirming the stated complaint.This device was manufactured in 2013, showing signs of wear from use.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history on the listed part revealed no prior complaints for the listed batch.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.No additional actions are being taken at this time; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 DEPTH GAUGE
Type of Device
PROSTH, HIP, SEMI-CONST, MET/CERAM/POLY, CEMENT OR NON-POROUS, UNCEMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8370515
MDR Text Key137157213
Report Number1020279-2019-00810
Device Sequence Number1
Product Code LZO
UDI-Device Identifier03596010599438
UDI-Public03596010599438
Combination Product (y/n)N
PMA/PMN Number
K093047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71364451
Initial Date Manufacturer Received 02/04/2019
Initial Date FDA Received02/26/2019
Supplement Dates Manufacturer Received02/04/2019
Supplement Dates FDA Received03/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-