• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH 5.5 EXP VERSE SCREW 7.0 X 45; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH 5.5 EXP VERSE SCREW 7.0 X 45; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 199721745
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during the a surgery on (b)(6) 2019, dr.(b)(6) became aware that the inner thread of screw was defective.The surgeon had to remove the screw and rod, because the locking screw could not screw in.A new screw has been inserted and a surgery delay of 30 min occured.
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4).Udi: (b)(4).Visual examination of the screw revealed that the threads of the tulip head were torn off completely.Some pieces of the torn thread of the tulip head were stuck in between the threads of correction key.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.With the information provided, a definitive root cause for the torn thread of the tulip head cannot be determined.Noted damage suggests that inadvertently cross threading of the correction key occurred upon insertion into the tulip head.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE SCREW 7.0 X 45
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key8370962
Report Number1526439-2019-51385
Device Sequence Number1203342
Product Code NKB
UDI-Device Identifier10705034442105
UDI-Public(01)10705034442105
Combination Product (Y/N)N
PMA/510(K) Number
K142185
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number199721745
Device Catalogue Number199721745
Device Lot Number102633
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2019
Initial Date Received by Manufacturer 01/31/2019
Supplement Date Received by Manufacturer03/11/2019
04/25/2019
Initial Report FDA Received Date02/26/2019
Supplement Report FDA Received Date03/29/2019
04/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-