• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.4MM VA-LCP 2-CLMN VLR DSTL RADIUS PL 6H HD/3H SHAFT/RIGHT; PLATE,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.4MM VA-LCP 2-CLMN VLR DSTL RADIUS PL 6H HD/3H SHAFT/RIGHT; PLATE,FIXATION,BONE Back to Search Results
Model Number 02.111.630
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is not expected to be returned for manufacturer review/investigation.Product was not returned.Device history records review could not be completed without lot number.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an (b)(6) 2019, surgeon was using a variable angle locking compression plate, while drilling the radial styloid screw with an unknown 1.8 drill bit and the variable angle guide.During the procedure, a 2.4 mm variable angle locking screw kept spinning and wouldn't engage into the plate.Surgeon then tried another screw of the same style but did not work.Surgeon tried a non variable angle locking screw, 1.8mm variable angle buttress pin, and 2.7 variable angel screw, but none of them would work.Surgeon then took a variable angle locking screw again and got it as tight as possible.It was still spinning but seemed to have enough purchase to stay in.The screw was left in the patient as essentially a cortical non locking screw.It had purchase in the bone so surgeon felt it was worth leaving in, but didn't have purchase in the plate.The plate remained in the patient.There was surgical delay of ten (10) minutes.Procedure outcome was successful competed.Patient outcome was successful.Concomitant medical products: drill bit (part unknown, lot unknown, quantity 1); variable angle guide (part unknown, lot unknown, quantity 1).This report is for one (1) 2.4mm va lcp plate.This is report 1 of 7 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device evaluated by mfr: the device was received, and the product evaluation is in progress.No conclusion can be drawn.Additionally, device history records review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not received and not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Concomitant devices reported: 1.8mm drill bit (part # 310.509 , lot unknown, quantity 1), variable angle guide (part unknown, lot unknown, quantity 1).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.4MM VA-LCP 2-CLMN VLR DSTL RADIUS PL 6H HD/3H SHAFT/RIGHT
Common Device Name
PLATE,FIXATION,BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8371379
Report Number2939274-2019-56677
Device Sequence Number13877030
Product Code HRS
UDI-Device Identifier10886982031990
UDI-Public(01)10886982031990
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K083694
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.111.630
Device Catalogue Number02.111.630
Was Device Available for Evaluation? No
Date Returned to Manufacturer02/25/2019
Initial Date Received by Manufacturer 02/01/2019
Supplement Date Received by Manufacturer02/25/2019
04/02/2019
Initial Report FDA Received Date02/26/2019
Supplement Report FDA Received Date03/25/2019
04/02/2019
Patient Sequence Number1
-
-