| Model Number |
265716 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
01/02/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon of the silicone foley catheter burst inside of the patient's bladder.Patient required surgery to remove the missing pieces.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon of the silicone foley catheter burst inside of the patient's bladder.Patient required surgery to remove the missing pieces.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon to burst.Valve type: use luer syringe.Do not use needle.Visually inspect the product for any imperfections or surface deterioration prior to use.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.".
|
| |
|