Model Number 175810
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the catheter was difficult to remove after the saline was withdrawn from the balloon.The catheter was removed by using a lubricant and manual manipulation.The patient voided after the removal without difficulty.Once removed, the nurse tried to fill the balloon again with saline and noticed the balloon was only inflating on one side.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿1.Occlude drainage tubing a minimum of 3 inches below the sampling port by kinking the tubing until urine is visible under the access site.2.Swab surface of bard® ez-lok® sampling port with antiseptic wipe.(fig.2) 3.Using aseptic technique, position the syringe in the center of the sampling port.Press the syringe firmly and twist gently to access the sampling port.(fig.3) 4.Slowly aspirate urine sample into syringe and remove syringe from sample port.5.Unkink tubing if necessary and transfer urine specimen into specimen cup or follow hospital protocol.Discard syringe according to hospital protocol.6.Follow established hospital protocol for specimen labeling and transport to lab.Sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not resterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the catheter was difficult to remove after the saline was withdrawn from the balloon.The catheter was removed by using a lubricant and manual manipulation.The patient voided after the removal without difficulty.Once removed, the nurse tried to fill the balloon again with saline and noticed the balloon was only inflating on one side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER
Common Device Name
LUBRISIL FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8371905
Report Number1018233-2019-01039
Device Sequence Number13910255
Product Code EZL
UDI-Device Identifier00801741034282
UDI-Public(01)00801741034282
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K984084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number175810
Device Catalogue Number175810
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/04/2019
Supplement Date Received by Manufacturer03/29/2019
Initial Report FDA Received Date02/26/2019
Supplement Report FDA Received Date04/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age9 YR
-
-