| Model Number |
MICL13.2 |
| Medical Device Problem Codes |
Contamination (1120); Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lens was returned in liquid in vial.Visual inspection found one haptic torn.No similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon inserted a 13.2mm micl13.2 implantable collamer lens, -16.0 diopter, in the patients right eye (od) and noted foreign material on the lens.The lens was removed.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The reporter indicated the surgeon inserted a 13.2mm micl13.2 implantable collamer lens, -16.0 diopter, into the patients right eye (od) and had an issue getting the lens (icl) to unfold.After trying a few techniques he felt it was best to remove the lens with no patient injury.The backup lens was implanted successfully.Date of report: 25-jan-2019.Date received by manufacturer: 25-jan-2019.Device code - 1254.Claim#: (b)(4).
|
| |
|