BARD ACCESS SYSTEMS GLIDEPATH LONG-TERM HEMODIALYSIS CATHETER; CATHETER, INTRAVASCULAR, THERAPEUTIC, LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 5397230
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that upon opening the chronic catheter kit, the white plastic tunneler tip was allegedly cracked.There was no reported patient contact.
 
Additional Manufacturer Narrative
Manufacturing review: a dhr review is not required, however it was requested.The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one 14.5 fr d/l glidepath 23 cm str hemodialysis catheter with surecuff kit was returned for evaluation.Visual and microscopic evaluations were performed.The investigation is confirmed for damaged/defective tunneler, as the residual proximal catheter attachment stub of (7.75 to 8.02) mm length and detached stub tip of (4.09 to 4.71) mm length to the arrow head base and 3.14 mm width at the fracture end.Further examination showed the fracture break profiles to match up.The definitive root cause for the reported damaged tunneler issue could not be determined based upon available information.It is unknown whether patient and/or procedural issues contributed to the event.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that upon opening the chronic catheter kit, the white plastic tunneler tip was allegedly cracked.There was no reported patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDEPATH LONG-TERM HEMODIALYSIS CATHETER
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC, LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8372404
Report Number3006260740-2019-00381
Device Sequence Number16418408
Product Code MSD
UDI-Device Identifier00801741012303
UDI-Public(01)00801741012303
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K051748
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Model Number5397230
Device Catalogue Number5397230
Device Lot NumberRECW0970
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2019
Initial Date Received by Manufacturer 01/30/2019
Supplement Date Received by Manufacturer03/27/2019
Initial Report FDA Received Date02/26/2019
Supplement Report FDA Received Date04/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-