• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and was unable to duplicate the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted physio-control to report that they were using this device on a patient and when they shocked the patient the device's display went completely black and the device's was not responding.There was a back up device available but it is unknown whether or not that device was used.As a result, defibrillation therapy may have been unavailable, if it was needed.There were no reports of any adverse effects to the patient as a result of the reported issue.Physio-control contacted the customer to request additional information on the patient.No response has been received from the customer.
 
Additional Manufacturer Narrative
Physio downloaded the device's electronic records from the event for review and verified the reported issue.Physio-control has determined that the likely cause of the reported issue was due to a field programmable gate array component (¿fpga¿) on the system pcb assembly.Timing issues in the fpga firmware may intermittently prevent a required system reboot following a defibrillator shock which results in the lockup issue.During the evaluation, the device lockup condition occurred once in every 800 to 1200 defibrillator discharge cycles.New fpga firmware has been created to resolve the timing issues that were identified.Validation testing of the new fpga firmware has confirmed that the updated firmware is effective in correcting intermittent device lockup issue following a defibrillator shock has been corrected.
 
Event or Problem Description
The customer contacted physio-control to report that they were using this device on a patient and when they shocked the patient the device's display went completely black and the device's was not responding.There was a back up device available but it is unknown whether or not that device was used.As a result, defibrillation therapy may have been unavailable, if it was needed.There were no reports of any adverse effects to the patient as a result of the reported issue.Physio-control contacted the customer to request additional information on the patient.No response has been received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key8372601
Report Number0003015876-2019-00324
Device Sequence Number1203163
Product Code MKJ
UDI-Device Identifier00883873911815
UDI-Public00883873911815
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001238
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2019
Initial Date Received by Manufacturer 02/04/2019
Supplement Date Received by Manufacturer04/01/2019
Initial Report FDA Received Date02/26/2019
Supplement Report FDA Received Date04/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3015876-12/28/2018-001-C
Patient Sequence Number1
-
-