| Model Number |
DDBB1D4 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to implant, the implantable cardioverter defibrillator (icd) interrogated a ¿gray bar¿ status indicating that the battery voltage level is not acceptable for implant.The device battery longevity was less than expected.The device was not used for implant.There were no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
| |
|