| Catalog Number |
110050100 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994); Not Applicable (3189); No Code Available (3191)
|
| Date of Event |
02/06/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.(b)(4).
|
| |
|
Event or Problem Description
|
|
Revision/removal of delta xtend reverse shoulder arthroplasty the patient experienced a post-op device malfunction during removal of prosthesis.The patient experienced an adverse event such as infection, non-union, allergic reaction, osteoporosis, overloading, pain, degenerative diseases, bleeding or oozing.The device was explanted due to pain and impingement which required revision surgery.It was suspected undersize of original implant that caused pain / impingement.The patient swabbed for infection due to painful movement.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|