Model Number 10602
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/04/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is a combination product.
 
Event or Problem Description
It was reported that stent fracture occurred.The severely stenosed target lesion was located in a mildly tortuous and moderately calcified circumflex artery as well as the left main artery.A 4.00 x 16 synergy stent was selected for use.During percutaneous coronary intervention, there were two balloons that were down and the stent expanded well.As the physician was removing the device from the circumflex artery, it was noted that the distal segment of the device broke probably about 60 cm from the distal end, and the broken piece remained inside the patient.Subsequently, the physician was able to take a 2.5 x 15 balloon past the synergy delivery device, trapped it and pulled the piece that was broken off back into the guide catheter and removed the whole system to retrieve the broken part.Furthermore, the broken part of the synergy was removed, the case was able to continue afterwards.The procedure was completed with the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8374886
Report Number2134265-2019-01402
Device Sequence Number16545840
Product Code NIQ
UDI-Device Identifier08714729840138
UDI-Public08714729840138
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2020
Device Model Number10602
Device Catalogue Number10602
Device Lot Number22618540
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2019
Initial Report FDA Received Date02/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-