Model Number 41-1000
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Communication received had an image of a floating set screw at distal end of construct.Model number was confirmed through case sheet provided.A lot number was not provided, so a device history record review and product investigation cannot be completed, as the device remains in-situ.Review of labeling notes: intra-operative warnings: if the system/construct contains screws, prior to soft tissue closure, recheck all screws to ensure they are tightened.Failure to do so may cause loosening of the other components.Postoperative warnings: the patient should be advised to limit and restrict physical activities, especially lifting and twisting motions and any type of sport participation.The patient should be advised that implants may bend, break or loosen despite restriction in activity.Possible adverse events: bending, disassembly or fracture of implant and components.Loosening of spinal fixation implants may occur due to inadequate initial fixation, latent infection, and/or premature loading, possibly resulting in bone erosion, migration or pain.It is unknown if the patient sustained a fall.The patient's anatomical condition and the effects on the stability of the implant and/or construct may be unknown contributing factors.The root cause of this event cannot be determined.No conclusions can be drawn.
 
Event or Problem Description
On (b)(6) 2018, the patient underwent posterior fixation surgery from levels t10-s1 using the mariner pedicle screw system.At the four week follow-up appointment, it was discovered that a set screw had disassociated from the screw tulip.The patient had subsequent follow-up appointments and radiographs were taken on (b)(6) 2018, (b)(6) 2018 and (b)(6) 2018.Patient is asymptomatic.No revision scheduled at this time.Surgeon will monitor patient progress.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MARINER PEDICLE SCREW SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
SEASPINE INC.
5770 armada drive
carlsbad CA 92008
MDR Report Key8375795
Report Number3012120772-2019-00006
Device Sequence Number1620391
Product Code NKB
UDI-Device Identifier10889981065065
UDI-Public10889981065065
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K160902
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number41-1000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2019
Initial Report FDA Received Date02/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
-
-