ENCORE MEDICAL L.P. RSP SHOULDER; RSP HUM SOCKET INSERT, 36MM +4MM, STANDARD Back to Search Results
Model Number 508-00-436
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/29/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery; due to the patient having dislocated.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient dislocated.The previous surgery and the revision detailed in this investigation occurred over 39 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.This investigation is limited in scope as limited information was provided to djo surgical - austin for examination and the part associated with this investigation was not returned to djo surgical for review.A review of the implant device history records (dhr) shows that the reported components used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to dislocation.There are many factors that may contribute to the event that are outside the control of djo surgical are patient activities, inadequate soft tissue support, excessive range of motion or trauma.Due to the short time between the original and revision surgery, the event may have possibly occurred due to lack of post-operative care or patient non-compliance with medical instructions.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's dislocation.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Common Device Name
RSP HUM SOCKET INSERT, 36MM +4MM, STANDARD
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8376167
Report Number1644408-2019-00156
Device Sequence Number14257719
Product Code KWS
UDI-Device Identifier00888912024686
UDI-Public(01)00888912024686
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K100741
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2019
Device Model Number508-00-436
Device Catalogue Number508-00-436
Device Lot Number919F1028
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/29/2019
Supplement Date Received by Manufacturer03/04/2019
Initial Report FDA Received Date02/27/2019
Supplement Report FDA Received Date03/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-