WILLIAM A. COOK AUSTRALIA, PTY LTD NATHANSON LIVER RETRACTOR EXTRA LARGE; GAD RETRACTOR - MURDOCH MECHANICAL ARM Back to Search Results
Catalog Number NLRS-1002
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Liver Damage/Dysfunction (1954); No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Event Description
This event was identified in the following paper: a novel liver retraction method in laparoscopic gastrectomy for gastric cancer (cea25022_revision 004).(b)(6) noted that the use of a nathanson retractor is an independent risk factor for elevated liver enzymes.
 
Event Description
This event was identified in the following paper: a novel liver retraction method in laparoscopic gastrectomy for gastric cancer (cea25022_revision 004).The department of surgery, internal medicine, research institute for covergence of biomedical science and technology, pusan national university yangsan-si hospital, korea noted that the use of a nathanson retractor is an independent risk factor for elevated liver enzymes.
 
Manufacturer Narrative
The device was not returned for evaluation.The lot number was not supplied, and therefore a review of the device history record could not be performed.The device ifu states: - step 4: under direct laparoscopic vision, manoeuvre the curved edge of the nathanson liver retractor in a superior direction under the left lobe of the liver.- step 5: continue placement of the ports.When this is complete, attach the nathanson liver retractor to the retractor arm and elevate the liver to the desired position.Periodic release of the retractor pressure may be necessary throughout the procedure to minimise the risk of hepatic ischaemia.Based on the information provided in it can be concluded that the adverse effect of the patient was most likely caused by a combination of patient-related factors, procedural-related factors, and insufficiently following the ifu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATHANSON LIVER RETRACTOR EXTRA LARGE
Type of Device
GAD RETRACTOR - MURDOCH MECHANICAL ARM
Manufacturer (Section D)
WILLIAM A. COOK AUSTRALIA, PTY LTD
95 brandl street
eight mile plains
brisbane QLD 4 113
AS  QLD 4113
MDR Report Key8376192
MDR Text Key137364385
Report Number9680654-2019-00007
Device Sequence Number1
Product Code GAD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type literature
Type of Report Initial,Followup
Report Date 03/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberNLRS-1002
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/15/2019
Initial Date FDA Received02/27/2019
Supplement Dates Manufacturer Received02/15/2019
Supplement Dates FDA Received03/12/2019
Patient Sequence Number1
Patient Outcome(s) Other;
-
-