• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM 4MAX REPERFUSION CATHETER; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM 4MAX REPERFUSION CATHETER; NRY Back to Search Results
Catalog Number 4MAXC
Medical Device Problem Codes Break (1069); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2019-00372.
 
Event or Problem Description
During preparation for a medical procedure, two penumbra system 4max reperfusion catheters (4maxcs) broke in two parts upon removal from the packaging.The damage to the two 4maxcs occurred prior to use and, therefore, were not used in the procedure.The procedure was completed using a new 4maxc.
 
Additional Manufacturer Narrative
Results: the first evaluated 4maxc was fractured approximately 53.5 cm from the hub.Conclusions: evaluation of the first and second returned 4maxcs revealed that both catheters were fractured.The first evaluated 4maxc was fractured in the proximal shaft.The second evaluated 4maxc was fractured underneath the strain relief.If the devices were forcefully retracted from their packaging at an extreme angle, damage such as a fracture may occur.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2019-00372.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM 4MAX REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8376383
Report Number3005168196-2019-00371
Device Sequence Number19364679
Product Code NRY
UDI-Device Identifier00814548012414
UDI-Public00814548012414
Combination Product (Y/N)Y
Initial Reporter CountryFR
PMA/510(K) Number
K160449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 01/29/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/22/2021
Device Catalogue Number4MAXC
Device Lot NumberF84795
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2019
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 01/29/2019
Supplement Date Received by Manufacturer04/04/2019
Initial Report FDA Received Date02/27/2019
Supplement Report FDA Received Date05/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-