| Catalog Number |
01.10.10.131 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
01/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 25 february 2019: lot 1111995c: (b)(4) items manufactured and released on 10-apr-2018.No anomalies found related to the problem.To date, no other similar event has been reported on this lot.
|
| |
|
Event or Problem Description
|
|
During the impaction the ball of the broach handle fell out from the passing hole and the spring was broken, no debris fell in the patient.There was a delay of max.Five minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection performed by r&d project manager: the spring ball plunger come apart.It is possible that this occurrence could have been influenced by instrument wear or variation in production/assembly, however this cannot be confirmed.
|
| |
|