C.R. BARD, INC. (COVINGTON) -1018233 BARD® ALL PURPOSE URETHRAL CATHETER Back to Search Results
Model Number 0094160
Medical Device Problem Code Misassembled (1398)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Date of Event 02/06/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the drainage hole on the catheter was more near the tip and the tip bent easily when given a little force.There was no patient injury.
 
Event or Problem Description
It was reported that the drainage hole on the catheter was more near the tip and the tip bent easily when given a little force.There was no patient injury.
 
Additional Manufacturer Narrative
The reported event was confirmed.Red rubber latex intermittent catheters were returned.A sampling plan was created and 20 samples of the lot of 100 were tested and two of the catheters were found to be out of specifications.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿caution: this product contains natural rubber latex which may cause allergic reactions.With funnel end 16¿ (40.6cm) length, two eyes, x-ray opaque rubber warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.For urological use only.U.S.Product.Packaged in mexico caution: federal (usa) law restricts this device to sale by or on the order of a physician.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® ALL PURPOSE URETHRAL CATHETER
Common Device Name
URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8377464
Report Number1018233-2019-01072
Device Sequence Number313038
Product Code KOD
UDI-Device Identifier00801741034039
UDI-Public(01)00801741034039
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2022
Device Model Number0094160
Device Catalogue Number0094160
Device Lot NumberNGCW1495
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2019
Initial Date Received by Manufacturer 02/06/2019
Supplement Date Received by Manufacturer06/03/2019
Initial Report FDA Received Date02/28/2019
Supplement Report FDA Received Date06/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-