• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-50
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/01/2019
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova deutschland manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective display panel.The technician replaced the panel to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The panel was requested back to livanova deutschland for further investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a centrifugal pump system with tubing clamp could not be controlled as the speed knob was blocked during procedure.There was no report of patient injury.
 
Event Description
See initial.
 
Manufacturer Narrative
H.10: the reported malfunction could be reproduced.It was not possible to turn the encoder, the shaft was stuck.Moreover, the knob and other accessories of the encoder were mechanically damaged.The cause of the event has been identified in a defective shaft angle encoder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8377685
MDR Text Key138815144
Report Number9611109-2019-00117
Device Sequence Number1
Product Code DWA
UDI-Device Identifier04033817900993
UDI-Public(01)04033817900993(11)181219
Combination Product (y/n)N
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-50
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-