| Model Number |
4323 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Pain (1994); No Code Available (3191)
|
| Date of Event |
02/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient described recent pain at the intercostal incision site.Ct ordered by implanting surgeon and possible breathing sensor migration into pleural space.Hardware/equipment is functioning as it should, patient ahi has been reduced from 47 to 1.4.
|
| |
|
Event or Problem Description
|
|
The patient complained of pain in the intercostal incision site.The issue was resolved by repositioning the sensor lead at revision on (b)(6) 2019.No new product was implanted, the originally implanted breathing sensor was repositioned.
|
| |
|
Search Alerts/Recalls
|