| Model Number |
27011L |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Codes |
Excessive Tear Production (2235); Patient Problem/Medical Problem (2688)
|
| Date of Event |
01/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
We evaluated the subject semi-rigid ureteroscope and we found the distal end is a little rough from long uses.Scopes shaft is bent and with approximately 60% broken light fibers.The seni-rigid ureteroscope was in use for approximately 12 years, purchased (b)(6) 2007.
|
| |
|
Event or Problem Description
|
|
Ureteroscope was placed into ureter and advanced toward the kidney.During removal, the ureter stuck to scope and was avulsed.The ureter could not be repaired and a nephrectomy was performed.
|
| |
|