• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEVICOR MEDICAL PRODUCTS, INC. MAMMOMARK BREAST BIOPSY SITE MARKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEVICOR MEDICAL PRODUCTS, INC. MAMMOMARK BREAST BIOPSY SITE MARKER Back to Search Results
Model Number MMK1002
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 08/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mammomark biopsy site identifier is a sterile, single use device intended for use after an open surgical or percutaneous breast biopsy procedure to mark the biopsy site.The device is not available for analysis, which precludes a full investigation and analysis of the root cause.However, this failure mode is identified in the risk management file and will occur at the end of the procedure and involves the marker tip physically breaking.The failure mode involves the tip of a side deploy marker shearing on the cutter of a disposable probe and separating from the body of the marker.Should this failure mode occur, there is a chance the broken marker tip may be left in the patient's breast.Medical intervention is required to remove the sheared tip.Based on patient consequences of unintended piece of the device of the device left in the biopsy site, and the additional surgical procedure to remove, and purusant to 21 cfr 803.We are submitting this medwatch report.
 
Event or Problem Description
Devicor medical products, inc.Received a user medwatch stating, "(b)(6) 2018 mammogram stereotactic right breast biopsy for indeterminate right breast microcalcifications, multiple passes were made under ultrasound guidance using a vacuum assisted 10-guage elite device.A titanium clip was placed at biopsy site.Specimen was sent to pathology for review.Microcalcifications were noted in the specimen radiograph.Post stereotactic biopsy right breast mammogram confirmed titanium clip was in the appropriate location.In addition, on the right mlo view, a 12x2.4 mm cylindrical structure, which likely corresponded to the radiopaque tip of the mammotome mark which sheared off during the procedure, located approx 7cm proximal to the titanium clip, superficially.The pt was informed of the sheared mammomark catheter tip and concurred with the surgeon's plan to leave the tip in place.The surgical pathology report revealed benign microcalcifications.The pt had no complaints of pain until (b)(6) 2018.On (b)(6) 2018, the pt underwent exploration and removal of foreign body in right breast under local anesthesia plus sedation.The tip of the mammotome was taken out and sent to pathology.On (b)(6) 2018 , the pt was seen in the surgery clinic and was doing well with well healing wounds.Dates of use: (b)(6) 2018.Diagnosis or reason for use: stereotactic right breast biopsy.The product was not compounded; the product was not over-the-counter".Report number mw5080699.This has been documented in our system as record # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAMMOMARK BREAST BIOPSY SITE MARKER
Common Device Name
BREAST BIOPSY SITE MARKER
Manufacturer (Section D)
DEVICOR MEDICAL PRODUCTS, INC.
300 e. business way
fifth floor
cincinnati OH 45241
MDR Report Key8379864
Report Number3008492462-2019-00011
Device Sequence Number19586544
Product Code NEU
UDI-Device Identifier00841911101628
UDI-Public(01)00841911101628(17)181201(10)F11721481D
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K082278
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/01/2018
Device Model NumberMMK1002
Device Catalogue NumberMMK1002
Device Lot NumberF11721481D
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2019
Initial Report FDA Received Date02/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-