• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS REFURB, SITE-RITE 8; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS REFURB, SITE-RITE 8; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770500R
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.Results are expected soon.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Per picc nurse: the unit was plugged in and had 78% power but just shutdown and would not reboot for more than 5 minutes.
 
Event or Problem Description
Per picc nurse: the unit was plugged in and had 78% power but just shutdown and would not reboot for more than 5 minutes.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of abrupt shutdown was confirmed.The scanner was charged to 100%, when the ac adaptor was removed the scanner ran on battery power for approximately 30 minutes.The root cause is an internal failure in the lithium-ion battery.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Per picc nurse: the unit was plugged in and had 78% power but just shutdown and would not reboot for more than 5 minutes.
 
Additional Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of abrupt shutdown was confirmed.The scanner was charged to 100%, when the ac adaptor was removed the scanner ran on battery power for approximately 30 minutes.The root cause is an internal failure in the lithium-ion battery.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURB, SITE-RITE 8
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8380507
Report Number3006260740-2019-00435
Device Sequence Number1651618
Product Code IYO
UDI-Device Identifier10801741104128
UDI-Public(01)10801741104128
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K152554
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Nurse
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9770500R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2019
Event Location Hospital
Initial Date Received by Manufacturer 02/09/2019
Supplement Date Received by Manufacturer05/03/2019
10/11/2019
Initial Report FDA Received Date02/28/2019
Supplement Report FDA Received Date05/21/2019
11/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2243072-10/11/2019-013-C
Patient Sequence Number1
-
-