| Model Number |
LO-PRO SCRW,TI,3.0MMX 28MMCORT |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
01/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was received but has not yet been evaluated.Once evaluated, a supplemental report will be submitted.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that two screws broke during a lisfranc procedure.The first screw ar-8740-52pts broke in half when spanning between the first metatarsal and the second metatarsal.Both broken pieces were removed from the patient.The second screw that broke was an ar-8933-28 which was being used in the plate.The head broke off the screw; the remainder of the screw remains in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint was confirmed.A fracture was observed on the underside of the screw.A critical dimensional measurement was done on the returned device.The device met all specifications as received.The most likely cause for this event to have occurred is by over torqueing the screw when inserting into the plate and hard bone.
|
| |
|