• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS PCA MODULE SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS PCA MODULE SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 10800173
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.Requested patient demographics however not provided by the customer.
 
Event or Problem Description
The customer reported the pca set was punctured.The event occurred in the oncology unit.Although requested there are no additional details provided at this time.There was no indication of patient harm.A noted received with the product documented; "date adm (b)(6) 2019".
 
Event or Problem Description
The customer reported the pca set was "punctured".The event occurred in the oncology unit.Although requested there are no additional details provided at this time.There was no indication of patient harm.A noted received with the product documented; "date adm (b)(6) 2019".
 
Additional Manufacturer Narrative
The customer¿s report the pca set was "punctured" was confirmed.Visual inspection of the set noted a cut in the suspect pca tubing approximately 2 inches below the antisiphon valve.There were no other anomalies observed.Functional testing confirmed leaking from the tubing approximately 2 inches below the antisiphon valve.Closer inspection of the leak under magnification observed a cut in the tubing that is approximately 0.151 in long.The root cause of the cut could not be identified.
 
Additional Manufacturer Narrative
Correction on initial report d.4: disregard lot # 11611800.
 
Event or Problem Description
The customer reported the pca set was "punctured".The event occurred in the oncology unit.Although requested there are no additional details provided at this time.There was no indication of patient harm.A noted received with the product documented ; "date adm (b)(6) 2019".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PCA MODULE SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key8382090
Report Number9616066-2019-00553
Device Sequence Number312457
Product Code FPA
UDI-Device Identifier10885403234385
UDI-Public10885403234385
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K960280
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10800173
Device Catalogue Number10800173
Device Lot Number11611800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2019
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date03/01/2019
Supplement Report FDA Received Date07/01/2019
07/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2420-0007, THERAPY DATE UNK; 2420-0007, THERAPY DATE UNK
-
-