| Model Number |
CRPLUS |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
A customer contacted physio-control to report that their device was showing all three icons (attention, charge-pak and service wrench).With all three icons illuminated, the device may not have sufficient power available to deliver effective defibrillation therapy to a patient, if it was needed.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control advised the customer that their device is not field serviceable and no longer under warranty.Physio recommended that the customer remove the device from service and a replacement device be obtained.The customer confirmed that the device was removed from service.The device has not been returned to physio-control for an evaluation.The cause of the reported event could not be determined.
|
| |
|
Event or Problem Description
|
|
A customer contacted physio-control to report that their device was showing all three icons (attention, charge-pak and service wrench).With all three icons illuminated, the device may not have sufficient power available to deliver effective defibrillation therapy to a patient, if it was needed.There was no patient use associated with the reported event.
|
| |
|