MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 9551
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/17/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.Device evaluated by mfr.: promus premier ous mr 3.00 x 24 mm stent delivery system was returned for analysis.A visual examination of the stent found that the stent was damaged on the 1st distal stent strut row with stent struts lifted and pulled distally.The undamaged stent od (outer diameter) was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of damage on the distal end of the stent.Tip damage most likely occurred due to device handling during preparation.A visual and tactile examination of the hypotube found multiple kinks along several locations of the hypotube shaft.This type of damage most likely occurred during handling of the device during preparation and/or prior to shipping for return.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 07feb2019.It was reported that device was defective.A 28x3.00mm and 24x3.00mm promus premier stent were selected for use.However, it was noted that both devices were defective.No known patient complications were reported.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8383485
MDR Text Key137692476
Report Number2134265-2019-01763
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/04/2019
Device Model Number9551
Device Catalogue Number9551
Device Lot Number0021481860
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-