The event was unconfirmed based on the evaluation.A red rubber intermittent catheter was returned for evaluation.The sample was received open and with original packaging.The sample was in good condition.The tip length was measured using calipers and was found to be within specifications.Tip length was related to eye size was found to be within specification and thus would not cause urine to drain more slowly (dimensional evaluation).The device was connected to a bag filled with water to simulate use.The device drained with no issues observed.A slip tip syringe was connected to the drainage funnel and used to flush the catheter.The water passed with no issues and no particulate was emitted from the drainage lumen (functional evaluation).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "instructions for use indications for use: for urological use only.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: this product contains natural rubber latex which may cause allergic reactions.Warnings: ¿ on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.¿ visually inspect the product for any imperfections or surface deterioration prior to use.¿ reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions urinary tract infection bleeding from the urethra irritation of the urethra instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands.".
|