| Model Number |
G17940 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # - exempt.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
| |
|
Event or Problem Description
|
|
It was reported during a ureterolithotripsy and double j procedure, the nforce nitinol helical stone extractor was tested prior to use.As reported, the handle showed resistance during testing.The basket closed but did not open anymore.This device was not used.A second nforce nitinol helical stone extractor from the same lot was tested.The basket didn´t close.The device was not used on the patient.Another extractor was opened and used to complete the procedure.There were no adverse effects to patient due to this occurrence.This report capturing the basket that did not close is related to mfr.Report # 1820334-2019-00536.
|
| |
|
Event or Problem Description
|
|
No new event information has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation/evaluation: a visual inspection and functional testing of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record, the instructions for use, and quality control data.One device was returned for investigation.The returned packaging confirms the reported complaint device lot number.The device was returned with the handle in the open position and the basket formation partially closed.The support sheath is bowed.The basket sheath is accordioned/ buckled 5 mm from the distal tip of the support sheath.Functional testing determined the handle actuates the basket formation to the fully open position, but does not retract the basket formation completely into the basket sheath.A review of the device history record found no non-conformances that would cause or contribute to the reported failure mode.A review of complaint history for the complaint device lot revealed one other complaint.The two complaints were determined to be unrelated.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: precaution: enclose the device in the sheath before removing from the tray/holder.Precaution: do not use excessive force to manipulate this device.Damage to the device may occur.Based on the investigation evaluation, there is no indication that a design or process related failure mode contributed to this event.Current controls for manufacturing are in place to assure functionality and device integrity prior to shipping.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.The returned device was found to have a basket that would not open / close.The basket sheath was found to be collapsed (buckled) near the blue support sheath.The buckled basket sheath was preventing the basket from opening and closing properly.All devices are inspected for damage and functionality before packaging.The cause for the collapsed sheath would have been a compressive force applied to sheath that was sufficient enough to cause the sheath to become damaged.It is possible the handle of the device was functioned with the sheath in a coiled position, that would cause a large compressive force on the sheath that could potentially cause buckling.There is no information related to device position when testing the basket function, so the cause for the complaint could not be conclusively determined.Per the quality engineering risk assessment, no further action is warranted.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Search Alerts/Recalls
|