| Model Number |
9734403 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.The instrument was returned for analysis.Functional testing found galling causing mating issue with the tracker.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that there was an issue with one of the site's navlock.The instrument was not accepting any instruments and also damaged two other instruments.It left grooves in the metal which prevents them from sliding into the shaft correctly.
|
| |
|