| Catalog Number |
10736401 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer stated that repeat testing was performed to confirm correct results which met the clinical picture.Siemens reviewed the in-house performance of the card lot and was found to be acceptable.The customer is advised to conduct a method comparison between epoc and the non-siemens lab instruments.As per the epoc user manual section 12.13.7 d, "blood samples must be well mixed in order to obtain accurate hematocrit results".The cause of this event is unknown.
|
| |
|
Event or Problem Description
|
|
The customer reported discrepant hemoglobin results on the epoc blood analysis system when compared to two non-siemens hematology lab analyzers.There was no report of injury due to this event.
|
| |
|