C.R. BARD, INC. (COVINGTON) -1018233 BARD® NASOGASTRIC SUMP TUBE; FILTERS FOR NG TUBE Back to Search Results
Model Number 0046000
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the markings on the ngt rubbed off and without the markings the nursing staff could not determine where the ngt was inserted or its location.Additionally, it was alleged that residue remained in the filters after use.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "caution: read all instructions prior to use.Indications for use: bard nasogastric sump tubes are intended to be used for: ¿ decompression of stomach by suction or aspiration of gastric contents.¿ short-term administration of term tube feeding, lavage fl uid and medications.Contraindications: ¿ patients with known tape or adhesive allergies.Warnings 1.Use with caution in patients with a history of head trauma, facial trauma, esophageal diseases and patients with potential for vomiting.2.Do not force nasogastric tube during insertions; damage to the nasal passage and mucosa and bleeding may occur.3.Measure insertion length carefully- excessive insertion length of tube into the stomach may lead to coiling and/or formation of tube-knot.4.Lubricate the tube generously with water soluble lubricant prior to insertion.Do not use petroleum-based products as they may be harmful to the respiratory tract.5.Refl ux of gastric contents into the blue vent lumen indicates that the suction lumen is obstructed or suction is too low.Routinely check for refl ux in the blue vent lumen and clear as per applicable directions.Failure to clear the obstruction or clear prevent® filter may cause gas and fluid buildup in stomach, aspiration of gastric contents, aspiration pneumonia and other complications.6.Do not inject fl uid through the prevent® filter as this may result in blockage and leakage of fi lter.7.Monitor patient for nasal erosion, sinusitis, esophagitis, esophagotracheal fi stula, gastric erosion and pulmonary & oral infections.Statlock® nasogastric stabilization device: avoid contact with alcohol or acetone; both can weaken bonding of components and statlock® stabilization device pad adherence.".
 
Event or Problem Description
It was reported that the markings on the ngt rubbed off and without the markings the nursing staff could not determine where the ngt was inserted or its location.Additionally, it was alleged that residue remained in the filters after use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® NASOGASTRIC SUMP TUBE
Common Device Name
FILTERS FOR NG TUBE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8384097
Report Number1018233-2019-01117
Device Sequence Number1432403
Product Code BSS
UDI-Device Identifier00801741052286
UDI-Public(01)00801741052286
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model Number0046000
Device Catalogue Number0042160
Device Lot NumberNGCU4332
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/08/2019
Supplement Date Received by Manufacturer04/01/2019
Initial Report FDA Received Date03/01/2019
Supplement Report FDA Received Date04/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-