• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; MONOFOCAL IOLS Back to Search Results
Model Number ZA9003
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted, give date: not applicable as lens was not implanted.If explanted, give date: not applicable as lens was not explanted.All pertinent information available to johnson and johnson surgical vision has been submitted.
 
Event or Problem Description
It was reported that a za9003 monofocal intraocular lens was damaged while inserting in the eye.Reportedly there was no patient injury and the procedure was completed without incident.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation ¿ yes, returned to manufacturer on 3/20/2019 device returned to manufacturer ¿ yes device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection at 10x microscope magnification showed the lens haptics distorted and one haptic was detached.The condition observed is consistent with a lens that has been prepared for surgical use.The reported issue was verified.However, there is no evidence to suggest that the complaint sample has been affected by the manufacturing process.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key8384169
Report Number2648035-2019-00265
Device Sequence Number19375886
Product Code HQL
UDI-Device Identifier05050474528956
UDI-Public(01)05050474528956(17)220719
Combination Product (Y/N)Y
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/09/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/19/2022
Device Model NumberZA9003
Device Catalogue NumberZA90030170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2019
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/05/2019
Supplement Date Received by Manufacturer04/04/2019
10/25/2020
Initial Report FDA Received Date03/01/2019
Supplement Report FDA Received Date05/02/2019
11/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Date Report Sent to FDA01/01/2005
-
-