Model Number E1452-6
Medical Device Problem Codes Arcing (2583); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, when the electrode needle was hooked on to the generator, the tip sparked and produced an open flame and had to be removed and replaced.The patient was on the table, but was not directly involved.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned electrode met specification.Visual inspection found that the tip of the electrode was charred with eschar.The date of expiration for the present lot is june 1, 2018, whereas the event date is jan 28, 2019.So, this is an expired lot.The investigation of the returned equipment did not identify anything that would have caused or contributed to the reported event.The investigation found that no components were missing.The customer did not report the generator settings.Visual inspection found that the electrode tip to be charred with eschar.The tip of the electrode discolors when activated and used on tissue.This is a normal part of electrosurgery and is not considered a defect.The investigation found the device to function normally and within specifications.The instructions for use (ifu) warns: the sparking and heating associated with electrosugery can provide an ignition source.Observe fire precautions at all times: when using electrosurgery in the same room with gases or flammable substances, prevent pooling of fluids and the accumulation of gases under surgical drapes or near the surgical site.Tissue buildup (eschar) on the tip of an active electrode poses a fire hazard, especially in oxygen-enriched environments, such as in throat or mouth procedures.Eschar plus high oxygen may create embers.Keep the electrode clean and free of all debris.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key8387361
Report Number1717344-2019-00273
Device Sequence Number1785039
Product Code GEI
UDI-Device Identifier10884521713161
UDI-Public10884521713161
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K955836
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2018
Device Model NumberE1452-6
Device Catalogue NumberE1452-6
Device Lot Number31650007X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2019
Initial Date Received by Manufacturer 02/07/2019
Supplement Date Received by Manufacturer07/30/2019
Initial Report FDA Received Date03/04/2019
Supplement Report FDA Received Date08/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-