| Model Number |
C20025 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: optia apheresis device; vortex implantable port; apheresis tubing set; large bore needle.(b)(4).The customer report that the extension tubing set pulled air was not confirmed.Visual inspection of the set showed dried blood residue around the area of the luer exit and specks of dried blood residue on both luer caps.Functional testing showed no anomalies.The root cause of the customer's report was not identified.The mating components used during the reported event were not received for investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported that they work in conjunction with the american red cross - western blood services division in providing apheresis therapy to their patients and that they are currently in the process of certifying their own apheresis unit.The customer reported that the optia device that performs apheresis treatment alarmed for "air detected in centrifuge" while using the extension tubing set.It was found that the extension tubing set pulled air from the outside into the apheresis kit and caused the kit/plasma exchange to malfunction leaving the nurse with the option of either performing a rinseback or discontinuing the apheresis therapy.Instead, the nurse replaced the extension tubing set to a different product and the issue was resolved.There was no patient harm.
|
| |
|