| Catalog Number |
72202962S |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during hip arthroscopy, the customer noticed the distal tip of the scope scratched.The procedure delayed more than 30 minutes and it is unknown if back-up was available.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation was performed by the supplier and could confirm the customer complaint for the distal tip was damaged.A visual inspection was performed and showed the scope to have distal tip damage and a loose prism.This damage is caused by contact with another source.No manufacturing related defects were observed.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection and functional testing could not be performed because the device in question has not been returned for evaluation.Thus, the complaint could not be verified and a root cause could not be determined with confidence.It is difficult to perform an accurate evaluation of a failure without having the failed product to look at.As such the complaint is being closed without conclusion.However, if the product is returned in the future the complaint can be reopened and evaluated.
|
| |
|