• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. VIDEOARTHORSCOPE HD 4MM X 70 DEG; NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. VIDEOARTHORSCOPE HD 4MM X 70 DEG; NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) Back to Search Results
Catalog Number 72202962S
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/15/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during hip arthroscopy, the customer noticed the distal tip of the scope scratched.The procedure delayed more than 30 minutes and it is unknown if back-up was available.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
An evaluation was performed by the supplier and could confirm the customer complaint for the distal tip was damaged.A visual inspection was performed and showed the scope to have distal tip damage and a loose prism.This damage is caused by contact with another source.No manufacturing related defects were observed.
 
Additional Manufacturer Narrative
Visual inspection and functional testing could not be performed because the device in question has not been returned for evaluation.Thus, the complaint could not be verified and a root cause could not be determined with confidence.It is difficult to perform an accurate evaluation of a failure without having the failed product to look at.As such the complaint is being closed without conclusion.However, if the product is returned in the future the complaint can be reopened and evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOARTHORSCOPE HD 4MM X 70 DEG
Common Device Name
NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key8388467
Report Number3003604053-2019-00025
Device Sequence Number1922560
Product Code EOB
UDI-Device Identifier00885554024210
UDI-Public00885554024210
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K043395
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72202962S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2019
Initial Date Received by Manufacturer 02/21/2019
Supplement Date Received by Manufacturer04/10/2019
07/02/2019
Initial Report FDA Received Date03/04/2019
Supplement Report FDA Received Date04/11/2019
07/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-