MAUDE Adverse Event Report: C.R. BARD, INC. (GFO) LUTONIX 035 DRUG COATED BALLOON PTA CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (GFO) LUTONIX 035 DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9004
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Reocclusion (1985); Claudication (2550)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.The lot history review could not be conducted because the lot number, although requested, was not available from the user facility.The dhr review also was not conducted because the facility did not provide the lot number.Conclusion: the actual samples were not returned for evaluation.The dhr review was not conducted because the facility did not provide the lot number.Based on the instructions for use (ifu), the occurrence of reocclusion and claudication are inherent risks of any pta procedure.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.
 
Event or Problem Description
It was reported through a literature article that during the index procedure, three lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion in the superficial femoral artery (sfa) and popliteal artery.Approximately 6 months after the index procedure, the patient was re-examined due to right calf claudication after walking 100 meters.A new duplex ultrasound (dus) revealed greater than 75% reocclusion just proximal to the eluvia stent which was placed due to a dissection occurring after the use of lutonix dcb's during the index procedure.The health care professional (hcp) presumed the reocclusion was a result of neointimal hyperplasia from balloon trauma, given that there was no clinical or biochemical evidence of an acute inflammatory flare of giant cell arteritis (gca).In the opinion of the investigator, the occurrence of reocclusion is possibly related to the lutonix dcb.The remainder of the sfa and stent appeared widely patent with no significant stenosis.An angioplasty procedure is planned involving the use of dcb's.The lutonix dcb's were discarded by the user facility and are not available for return.No adverse patient outcomes were reported.This is one of three products involved with the reported event and the associated manufacturer¿s report numbers are 3006513822-2019-00034 and 3006513822-2019-00035.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
MDR Report Key8388658
Report Number3006513822-2019-00036
Device Sequence Number19390026
Product Code ONU
UDI-Device Identifier00801741088759
UDI-Public(01)00801741088759
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,literatur
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number9004
Device Catalogue Number9040413500150
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2019
Initial Report FDA Received Date03/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
-
-