| Model Number |
129416 |
| Health Effect - Clinical Codes |
Patient Problem/Medical Problem (2688); No Information (3190)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon of the bardex ic temp sensing foley catheter would not deflate.An ultrasound was required to help guide an 18g spinal needle through the perineum to pop the balloon.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.".
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon of the bardex ic temp sensing foley catheter would not deflate.An ultrasound was required to help guide an 18g spinal needle through the perineum to pop the balloon.
|
| |
|