• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KANEKA CORPORATION METACROSS RX PTA BALLOON DILATATION CATHETER; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KANEKA CORPORATION METACROSS RX PTA BALLOON DILATATION CATHETER; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number BD-L70200CR
Medical Device Problem Codes Material Separation (1562); Use of Device Problem (1670); Device Handling Problem (3265)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
Date of Event 02/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device history records (dhr) of the device concerned was reviewed: the production lot, to which the device concerned belongs, passed all in-process inspections for every product, and the finished product inspections on representative samples based on sampling plan.No nonconformity or abnormality in the manufacturing processes of the device concerned was found.The actual device concerned was returned and investigated: the proximal end of the catheter was cut.The balloon was broken at 40mm from the proximal shoulder of the balloon.The inner shaft and outer shaft were elongated or cut.The distal radiopaque marker was missing.Probable cause(s) and our comment: the balloon should have been injured by an edge of the stent strut and was ruptured during being inflated.In order to deflate the balloon that had inflated, the concerned device was retract into the child catheter, so the balloon was caught at the breakage part on the tip of the child catheter and was broken.The broken fragment of the balloon remained inside the blood vessel.No non-conformity or abnormality in the manufacturing processes of the device concerned was found, and accordingly, we determine that the event reported was caused by not any defect of the device but the user's handling issues.
 
Event or Problem Description
In-stent restenosis (occlusion) lesion.Using a guiding sheath and a 0.014 inch guide wire, after the catheter passed through the lesion the balloon was dilated, but it was ruptured at 14 atm.The child catheter (the catheter thinner than the guiding sheath) could be placed in the guiding sheath.Because the balloon of the catheter was a little inflated, it attempted to retract into the child catheter to deflate the balloon, but it could not be done.Ultimately, the distal end of the balloon broke, and the broken fragment remained in the patient's vessel.The fragment was fixed to the vessel wall with the stent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METACROSS RX PTA BALLOON DILATATION CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
KANEKA CORPORATION
2-3-18
nakanoshima, kita-ku
osaka-city, osaka 53082 88
JA  5308288
MDR Report Key8388929
Report Number3002808904-2019-00003
Device Sequence Number13957040
Product Code DQY
UDI-Device Identifier04540778153110
UDI-Public04540778153110
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K150865
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Physician Assistant
Type of Report Initial
Report Date (Section B) 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2019
Device Catalogue NumberBD-L70200CR
Device Lot NumberSP046435
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2019
Initial Report FDA Received Date03/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-