• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR MMT-7020A ENLITE3 5PK US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR MMT-7020A ENLITE3 5PK US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-7020A
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Calibration Problem (2890)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 170 mg/dl and the sensor glucose was 40 mg/dl.The customer was assisted with troubleshooting.Insulin delivery was suspended due to sensor values.Customer states the sensor value that triggered the suspend event was 70.Suspend on low limit in sensor settings was 40.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable rang.Customer reports they did not receive a sensor updating or sensor glucose not available alert.Customer reports they did receive a calibration not accepted alert.Customer did not report consecutive sensor alerts with different sensors.The device will be returned for analysis.
 
Additional Manufacturer Narrative
Inspected one opened/used sensor and found cannula retracted inside sensor base.Unable to confirm customer received sensor in said condition due to customer returned opened/used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR MMT-7020A ENLITE3 5PK US
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8389524
Report Number3004209178-2019-62581
Device Sequence Number19278487
Product Code OZP
UDI-Device Identifier20763000118864
UDI-Public(01)20763000118864(17)190409(10)HG2FRGH
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/09/2019
Device Model NumberMMT-7020A
Device Catalogue NumberMMT-7020A
Device Lot NumberHG2FRGH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2019
Initial Date Received by Manufacturer 02/25/2019
Supplement Date Received by Manufacturer02/12/2020
Initial Report FDA Received Date03/05/2019
Supplement Report FDA Received Date02/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age45 YR
Patient Weight118
-
-