| Model Number |
MMT-7020A |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 170 mg/dl and the sensor glucose was 40 mg/dl.The customer was assisted with troubleshooting.Insulin delivery was suspended due to sensor values.Customer states the sensor value that triggered the suspend event was 70.Suspend on low limit in sensor settings was 40.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable rang.Customer reports they did not receive a sensor updating or sensor glucose not available alert.Customer reports they did receive a calibration not accepted alert.Customer did not report consecutive sensor alerts with different sensors.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened/used sensor and found cannula retracted inside sensor base.Unable to confirm customer received sensor in said condition due to customer returned opened/used.
|
| |
|