| Catalog Number |
386807 |
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of the cathena 24gx0.75in straight bc the catheter anti reflex valve does not function after insertion or connection/disconnection and there was leaking past the valve and out of the hub.There were 2 occurrences.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of the cathena 24gx0.75in straight bc the catheter anti reflex valve does not function after insertion or connection/disconnection and there was leaking past the valve and out of the hub.There were 2 occurrences.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: 5 representative samples (unopened) were returned for investigation.The samples were subjected to blood escape time (bet) test.The samples passed the bet test, no leakage was observed.A review of past 12 months quality notification was performed and no quality notification was raised on the reported defect of leakage.No abnormality was detected during the production of the reported batch.The returned sample passed the bet test.Hence root cause could not be determined.However, there was a capa raised for blood droplet formation at the luer end of the catheter adapter (b)(4).Small id catheter tubing and small septum vents have been implemented to enhance the blood control function.
|
| |
|