• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSPIRE MEDICAL SYSTEMS INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSPIRE MEDICAL SYSTEMS INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD Back to Search Results
Model Number 4323
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Infiltration into Tissue (1931)
Date of Event 03/01/2019
Type of Reportable Event Serious Injury
Event or Problem Description
During implant, the sensing lead penetrated the space between the internal intercostals and lung.The issue was recognized, the area was oversewn with two layers of suture.The area was flooded with saline, anesthesia performed valsalva without any bubbles being seen.The lead was then moved to a different rib space successfully.Post-op chest x-ray did not show pneumothorax.The patient was clinically asymptomatic and discharged per usual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD
Common Device Name
INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD
Manufacturer (Section D)
INSPIRE MEDICAL SYSTEMS
5500 wayzata blvd.
suite 1600
golden valley MN 55416
MDR Report Key8391027
Report Number3007666314-2019-00023
Device Sequence Number19190588
Product Code MNQ
UDI-Device Identifier10855728005496
UDI-Public10855728005496
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P130008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/05/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number4323
Device Catalogue Number900-002-009
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2019
Initial Report FDA Received Date03/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Date Report Sent to FDA01/01/2005
-
-