EDWARDS LIFESCIENCES, PR SWAN-GANZ CCO/VIPTHERMODILUTION CATHETER; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 139F75
Device Problems Leak/Splash (1354); Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/19/2019
Event Type  malfunction  
Manufacturer Narrative
Our product evaluation laboratory received one model 139f75 swan/ganz catheter with a syringe attached.The balloon was found to be ruptured.The ruptured edges did not appear to match up.All through lumens were patent without any leakage or occlusion.No other visible damage or abnormality was observed from the catheter body or the returned syringe.A device history record review was completed and documented that the device met all specifications upon distribution.The customer report of a balloon issue was confirmed on evaluation.An engineering evaluation has been initiated to assess for any manufacturing-related processes which could be correlated to the complaint.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is standard practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.When there is separation of the balloon or fragments from the pulmonary artery catheter, the retained fragment will embolize to the lungs.Due to the large surface area of the pulmonary vasculature, this is generally well tolerated, but can lead to complications such as infection or small infarction.Pulmonary complications may result from improper inflation technique.To avoid damage to the pulmonary artery and possible balloon rupture, the balloon should not be inflated above the recommended volume.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that there was leakage from the balloon on a swan-ganz catheter.It is unknown if the leakage was discovered before or during use.There was no patient injury.Patient demographics were requested and not provided.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCO/VIPTHERMODILUTION CATHETER
Type of Device
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key8391282
MDR Text Key138359786
Report Number2015691-2019-00728
Device Sequence Number1
Product Code KRA
Combination Product (y/n)N
PMA/PMN Number
K924661
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/26/2020
Device Model Number139F75
Device Catalogue Number139F75
Device Lot Number61524820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2019
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-