MAUDE Adverse Event Report: ATRIUM MEDICAL OCEAN DRAIN; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRIUM MEDICAL OCEAN DRAIN; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number 2002-300
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/26/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
We are in the process of performing the investigation and will submit the follow-up report once the evaluation is completed.
 
Event or Problem Description
Report received stated that the drainage is unable to reach the 200ml graduation mark on the drain.
 
Additional Manufacturer Narrative
Analysis: the drain in question was disinfected prior to testing as it was still full of pleural fluid.Upon completion of the disinfection a 200ml beaker of blue colored water was poured into the first chamber of the drains through the patient tube line.The drain filled to the 200ml line on the chamber as expected.There were no signs that the fluid was leaking over into the other chamber at the 180ml fill mark as described in the complaint details.A review of the device history records indicates that this lot of chest drains passed all quality requirements.Conclusion: based on the results of the investigation atrium medical corporation cannot conclude that there was any fault of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCEAN DRAIN
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
ATRIUM MEDICAL
40 continental blvd
merrimack NH 03054
MDR Report Key8393060
Report Number3011175548-2019-00238
Device Sequence Number19564740
Product Code KDQ
UDI-Device Identifier00650862100105
UDI-Public00650862100105
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K043582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/13/2021
Device Model Number2002-300
Device Catalogue Number2002-300
Device Lot Number428202
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2019
Initial Date Received by Manufacturer 02/27/2019
Supplement Date Received by Manufacturer03/26/2019
Initial Report FDA Received Date03/05/2019
Supplement Report FDA Received Date03/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight2
-
-