| Model Number |
2002-300 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/26/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
We are in the process of performing the investigation and will submit the follow-up report once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
Report received stated that the drainage is unable to reach the 200ml graduation mark on the drain.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis: the drain in question was disinfected prior to testing as it was still full of pleural fluid.Upon completion of the disinfection a 200ml beaker of blue colored water was poured into the first chamber of the drains through the patient tube line.The drain filled to the 200ml line on the chamber as expected.There were no signs that the fluid was leaking over into the other chamber at the 180ml fill mark as described in the complaint details.A review of the device history records indicates that this lot of chest drains passed all quality requirements.Conclusion: based on the results of the investigation atrium medical corporation cannot conclude that there was any fault of the device.
|
| |
|