(b)(4).Used needle-suture piece and empty labeled winding former of product code y305, lot mlz109 were returned for analysis.During the visual inspection of opened samples (needle/suture pieces), the swage and attachment area were noted to be as expected.However, marks on the body of the needle and the end of the suture pieces cut that appears to be by the use of a surgical instrument could be observed.The other section of the suture was not returned for analysis and due to the condition of the samples the functional test can¿t be performed.Due to condition of the opened sample, the assignable cause of performance breakage suture suggests an improper handling.Used needle-suture piece and empty labeled winding former of product code y305, lot mlz109 were returned for analysis.During the visual inspection of opened samples (needle/suture pieces), the swage and attachment area were noted to be as expected.However, marks on the body of the needles and the end of the suture pieces cut that appears to be by the use of a surgical instrument could be observed.Also a knot was found near to the end of the strand.The other section of the suture was not returned for analysis and due to the condition of the samples the functional test can¿t be performed.Due to condition of the opened sample, the assignable cause of performance breakage suture suggests an improper handling.Labeled winding former with a suture piece each and one empty opened foil of product code y305, lot mlz109 were returned for analysis.During the visual inspection of suture piece, the end of the strand was cut that appears to be by the use of a surgical instrument could be observed.The other section of the suture was not returned for analysis.A functional test was performed and the tensile strength force was above the minimum requirement.Due to condition of the opened sample, the assignable cause of performance breakage suture suggests an improper handling.
|